Computer Body Posture Examination - construction

 

The apparatus takes the advantages of photogrammetric method as well as Moire effect. It is designed for a completely safe, non-invasive professional diagnostics of body posture. It makes possible to determine over 50 parameters and graphical records.  It is especially useful for screening examinations because measuring one person takes 2-3 minutes.

Budowa

 

1.      Power outlet with standard computer cable.

2.      Socket (9 openings) to connect apparatus with computer.

3.      Main fuse socket (Attention! The fuse can only be changed with off-line feeding cable).

4.      Main power switch.

5.      Fan opening (Attention! Absolutely you cannot allow a foreign body to get within the opening! On operating the device there is a spinning unit! You must not cover the opening!).

5b – ventilation gaps – they must not be covered during device operation!

6.      Data plate.

7.      Patient light.

8.      Camera.

9.      Visiting plate.

10.    Level setting for lateral inclination.

11.    Level setting for forward – backward inclination.

12.    Casing assembling bolts.

13.    Assembling bolts to assemble casing back to the base.

14.    Assembling bolts to assemble casing to the base.

15.    Apparatus front feet - adjustable.

Apparatus back feet - stationary.

 

 

PERFORMANCE CHARACTERISTICS

 Parameters

Measuring time:                 0.03s

Speed:                               app. 3 measurements per second

Average time of one measurement: app. 1 min

Working voltage:               230V, 50Hz

Power rating:                     160W / 60W  (depending on specific unit)

Lightning:                           halogenous bulb 24V-150W / 12V-50W (depending on the unit)

Dimensions:                                  300x420x120 / 250x370x110 (portable version)

Weight:                               app. 6kg

Examination distance:      2.6m / 2.0m (values for portable version)

Computer requirements: PENTIUM 1500; RAM.512 MB; 80 MB free on HDD;

Graphic board                   (1024x768, High Colour)

Operating system:                         WINDOWS XP

 

Service conditions:      

1.      The appliance should be used appropriately.

2.      All co-operative devices of the set have to be supplied from the same phase and have common zeroing.

3.      Power outlet 230V/50Hz has to have earth. It is recommended to supply through a circuit-breaker strip (e.g. ACAR). In case of oscillations or voltage drop it is necessary to apply an additional stabilizing device.  

4.      Apparatus has to work in rooms at the temperatures from +10°C till +25°C and relative humidity up to 80% at the temperature of +25°C.

5.      Apparatus should not be subject to vibrations, be operated next to strong electromagnetic fields and should be protected from an excessive sun exposure.

6.      It should be operated far from dust.